Department of Chemistry, Avinashilingam Deemed University for Women Coimbatore India
A heterogeneous precipitate have been used as ion carriers for the preparation of Na(I) selective membrane sensor. This electrode give near-nernstian responses in the linear concentration range of 1M to 1×10-3 M with detection limits of the order of 10-3 M. The stable potentiometric signals are obtained with in a short time period of 3 minutes.The effect of pH and the effect of medium have been studied found to give a better response.Selectivity coefficient values (log KpotNa,M) have been evaluated using fixed interference method.The sensor have also been used as indicator electrode in commercial products.
Copy the following to cite this article:
J. THAMARAI SELVI , "A new simple method for the preparation of sodium ion selective electrode-its preparation and analytical application ", Journal of Ultra Scientist of Physical Sciences, Volume 23, Issue 2, Page Number 445-448, 2016Copy the following to cite this URL:
J. THAMARAI SELVI , "A new simple method for the preparation of sodium ion selective electrode-its preparation and analytical application ", Journal of Ultra Scientist of Physical Sciences, Volume 23, Issue 2, Page Number 445-448, 2016Available from: http://ultraphysicalsciences.org/paper/634/